Validation and Testing

AI-Optimized Intravenous Therapy Control System — v4.2.0
Status: Pre-Clinical Research Reference
Language: C++17

Important: This document applies to a pre-clinical research system. It does not constitute clinical validation, regulatory evidence, or a claim of medical device approval.


Table of Contents

  1. Purpose
  2. System Under Test
  3. Build Verification
  4. Test Suite
  5. Test Results Summary
  6. Safety Invariant Analysis
  7. Configuration Constants and Bounds
  8. Failure Mode Analysis
  9. Known Gaps and Limitations
  10. Path to Clinical Validation

1. Purpose

This document provides a structured analysis, test documentation, and validation summary for the AI-Optimized Intravenous Therapy Control System. It covers:

  • Build verification across all supported targets
  • Documented test cases with expected and actual results
  • Safety invariant verification
  • Configuration constant review
  • Honest assessment of coverage gaps

This document is designed to:

  • Support research review and collaboration
  • Enable safety-critical code inspection
  • Provide a transparent baseline for future regulatory work

2. System Under Test

Module Inventory

Module Source File(s) Role
StateEstimator src/StateEstimator.cpp / .hpp Signal fusion, coherence scoring, energy proxy, cardiac reserve, risk scoring
AdaptiveController src/AdaptiveController.cpp / .hpp Demand modeling, coherence modulation, cardiac limiting, predictive boost
SafetyMonitor src/SafetyMonitor.cpp / .hpp Volume limits, cardiac load, rate-of-change, emergency overrides
SystemLogger src/SystemLogger.cpp / .hpp Structured NDJSON alert events, telemetry CSV, control CSV
NeuralStateEstimator src/NeuralStateEstimator.hpp 241-parameter feedforward network (optional, frugally-deep)
SimulationEngine src/simulation_engine.cpp / .hpp Parameterized physiological scenario driver
RestApiServer src/rest_api_server.cpp / .hpp Read-only HTTP API (optional, -DENABLE_REST_API)

Data Contracts

All inter-module data is exchanged through stable types defined in src/iv_system_types.hpp:

Type Purpose
Telemetry Raw sensor snapshot (hydration, HR, SpO₂, temp, lactate, fatigue, signal quality)
PatientProfile Static patient parameters (weight, age, conditions, safety limits)
PatientState Derived physiological state (coherence, energy proxy, cardiac reserve, risk score)
ControlOutput Infusion command with confidence and rationale string
AlertEvent Structured NDJSON alert with severity, code, source, message

Centralized Configuration

All tunable parameters are defined in src/config_defaults.hpp. No magic numbers appear in control or safety logic.


3. Build Verification

Targets

Target Command Description
Core binary make all Standard build, strict warnings
Test binaries make test Compiles and runs test_safety_monitor + test_state_estimator
Neural-enabled binary make neural Requires libfdeep-dev, libeigen3-dev
Neural tests make test_all Includes test_neural_estimator
Clean make clean Removes all build artifacts

Compiler Flags

g++ -std=c++17 -O2 -Wall -Wextra -pthread

All three flags (-Wall, -Wextra, and the C++17 standard) are enforced. The build produces zero warnings on the standard target.

Build Results (v4.2.0)

Target Result
make all ✅ Pass — zero warnings
make test ✅ Pass — all unit tests pass
make neural Requires optional libfdeep-dev / libeigen3-dev apt packages

4. Test Suite

Tests are located in the tests/ directory and compiled via make test or make test_all. Each test binary exits with code 0 on success and calls exit(1) on any assertion failure, making failures unambiguous in CI.

4.1 SafetyMonitor Tests

File: tests/test_safety_monitor.cpp
Binary: ./test_safety_monitor

Test: test_volume_limit

Verifies that the SafetyMonitor reduces max_allowed_rate and emits a VOLUME_LIMIT_APPROACH warning when cumulative infused volume exceeds 90% of the patient's 24-hour limit.

Parameter Value
Patient weight 70 kg
Cardiac condition false
Renal impairment false
Max safe infusion rate 1.5 ml/min
24-hour volume limit 70 × 35 = 2450 ml
90% warning threshold 2205 ml
Pre-loaded volume 2300 ml (via update_volume(100.0, 23.0))

Expected behaviour:

  • max_allowed_rate reduced to ≤ 0.3 ml/min
  • warnings string contains VOLUME_LIMIT_APPROACH

Result: ✅ Pass


Test: test_cardiac_reserve

Verifies that the SafetyMonitor reduces max_allowed_rate and emits a LOW_CARDIAC_RESERVE warning when the patient's cardiac reserve falls below the minimum threshold (0.2).

Parameter Value
Patient weight 70 kg
Cardiac condition false
Max safe infusion rate 1.5 ml/min
Injected cardiac reserve 0.1 (critically low)

Expected behaviour:

  • max_allowed_rate reduced to ≤ 0.5 ml/min
  • warnings string contains LOW_CARDIAC_RESERVE

Result: ✅ Pass


4.2 StateEstimator Tests

File: tests/test_state_estimator.cpp
Binary: ./test_state_estimator

Test: test_estimate_basic

Verifies that StateEstimator::estimate correctly passes through hydration and heart rate from a telemetry snapshot, and that the uncertainty estimate is reasonable for a healthy, high signal-quality input.

Input Value
hydration_pct 80.0 %
heart_rate_bpm 75.0 bpm
temp_celsius 37.0 °C
spo2_pct 98.0 %
signal_quality 1.0 (perfect)
Patient weight 70 kg
Baseline HR 70 bpm

Expected behaviour:

  • state.hydration_pct within 0.1 of 80.0
  • state.heart_rate_bpm equals 75.0
  • state.uncertainty < 0.5 (low uncertainty under perfect signal quality)

Result: ✅ Pass


4.3 Neural Estimator Tests

File: tests/test_neural_estimator.cpp
Binary: ./test_neural_estimator
Requires: -DENABLE_NEURAL_ESTIMATOR, libfdeep-dev, libeigen3-dev,
models/sensor_fusion_fdeep.json

The neural estimator is a 241-parameter feedforward network (Dense-16-ReLU → Dense-8-ReLU → Dense-1-Sigmoid) trained via knowledge distillation from the hand-crafted rule-based formula.

Test: test_load_and_healthy_patient

Loads the frugally-deep JSON model and runs inference for a healthy patient.

Input (normalised) Value
hydration_norm 0.80 (80%)
hr_norm 0.375 (75 bpm / 200)
spo2_norm 0.98
lactate_norm 0.10 (2 mmol/L / 20)
fatigue 0.30

Expected: E_T > 0.6 and ≤ 1.0


Test: test_stressed_patient

Runs inference for a severely stressed patient.

Input (normalised) Value
hydration_norm 0.40 (40%)
hr_norm 0.65 (130 bpm / 200)
spo2_norm 0.84
lactate_norm 0.40 (8 mmol/L / 20)
fatigue 0.90

Expected: E_T < 0.5 and ≥ 0.0


Test: test_rule_formula_agreement

Verifies that the neural model agrees with the reference rule-based formula to within a mean absolute error (MAE) of 0.08 across three representative spot-check samples. This threshold matches the training target in tools/train_sensor_fusion_model.py.

Sample hydration HR SpO₂ lactate fatigue Reference E_T
0 70% 70 bpm 98% 2 mmol/L 0.20 ~0.85
1 50% 100 bpm 92% 5 mmol/L 0.60 ~0.55
2 90% 80 bpm 99% 1 mmol/L 0.10 ~0.88

Expected: MAE across all three samples < 0.08

These tests require the optional frugally-deep runtime and are run via make test_all. The standard CI gate (make test) covers only the deterministic (non-neural) tests.


5. Test Results Summary

Test Module Result
test_volume_limit SafetyMonitor ✅ Pass
test_cardiac_reserve SafetyMonitor ✅ Pass
test_estimate_basic StateEstimator ✅ Pass
test_load_and_healthy_patient NeuralStateEstimator ✅ Pass (optional)
test_stressed_patient NeuralStateEstimator ✅ Pass (optional)
test_rule_formula_agreement NeuralStateEstimator ✅ Pass (optional)

All tests in the standard suite (make test) pass with exit code 0.


6. Safety Invariant Analysis

The following invariants are enforced by design. Each is traceable to source code.

Invariant 1 — Infusion Rate Bounds

Source: src/config_defaults.hpp, src/AdaptiveController.cpp, src/SafetyMonitor.cpp

MIN_INFUSION_RATE_ML_MIN = 0.1 ml/min
MAX_INFUSION_RATE_ML_MIN = 1.5 ml/min

AdaptiveController::decide clamps desired_rate to [0.1, profile.max_safe_infusion_rate] before forwarding to SafetyMonitor::evaluate. The safety monitor then independently caps the rate via max_allowed_rate. The final command can never exceed the lower of these two bounds.

Verified by: test_volume_limit, test_cardiac_reserve


Invariant 2 — Safety Monitor Cannot Be Bypassed

Source: src/AdaptiveController.cpp (lines 82–88)

auto safety_check = safety.evaluate(desired_rate, state, dt_minutes);
if (desired_rate > safety_check.max_allowed_rate) {
    desired_rate = safety_check.max_allowed_rate;
    safety_limited = true;
}

The controller always passes its computed rate through SafetyMonitor::evaluate and accepts any reduction imposed by it. There is no code path that skips the safety check.


Invariant 3 — Cardiac Reserve Limiting

Source: src/SafetyMonitor.cpp

MIN_CARDIAC_RESERVE = 0.2

When state.cardiac_reserve < 0.2, max_allowed_rate is capped at 0.5 ml/min and LOW_CARDIAC_RESERVE is appended to the warning string.

AdaptiveController::apply_cardiac_limiting provides a second, independent cardiac scaling step that reduces the rate before it reaches the safety monitor, providing defence in depth.

Verified by: test_cardiac_reserve


Invariant 4 — Volume Overload Protection

Source: src/SafetyMonitor.cpp

DAILY_VOLUME_PER_KG_ML = 35.0 ml/kg

Maximum 24-hour volume = weight_kg × 35.0 ml, further reduced by:

  • 30% for cardiac conditions (× 0.7)
  • 40% for renal impairment (× 0.6)

When projected volume exceeds 90% of this limit, max_allowed_rate is capped at 0.3 ml/min.

Verified by: test_volume_limit


Invariant 5 — Rate-of-Change Limiting

Source: src/SafetyMonitor.cpp, src/config_defaults.hpp

MAX_RATE_CHANGE_ML_MIN = 0.3 ml/min per control cycle

If the requested rate differs from the most recent rate by more than 0.3 ml/min, the rate is clamped to last_rate ± 0.3. This prevents step changes and oscillation in the infusion trajectory.


Invariant 6 — High-Risk State Throttling

Source: src/SafetyMonitor.cpp

HIGH_RISK_THRESHOLD = 0.75

When state.risk_score > 0.75, max_allowed_rate is capped at 0.6 ml/min and HIGH_RISK_STATE is flagged.


Invariant 7 — Tachycardia Detection

Source: src/SafetyMonitor.cpp

When state.heart_rate_bpm > profile.baseline_hr_bpm × 1.4, max_allowed_rate is reduced to 0.4 ml/min and TACHYCARDIA_DETECTED is flagged.


Invariant 8 — Emergency Minimum Rate

Source: src/SafetyMonitor.cpp

if (result.max_allowed_rate < 0.1 && state.hydration_pct < 50.0) {
    result.max_allowed_rate = 0.1;
    warnings << "EMERGENCY_MIN_RATE ";
}

If all safety checks together would reduce the rate below 0.1 ml/min, but the patient is critically dehydrated (hydration < 50%), the rate is forced back to the minimum safe level. This prevents total infusion cessation in critical cases.


Invariant 9 — Determinism

Source: src/StateEstimator.cpp, src/AdaptiveController.cpp

All control logic is deterministic. Given identical Telemetry and PatientProfile inputs, the system produces identical ControlOutput. No random number generators, hardware timestamps, or non-deterministic operations appear in the control path.


Invariant 10 — Signal Coherence Damping

Source: src/StateEstimator.cpp

calculate_coherence reduces the coherence score when:

  • HR is outside 40–180 bpm (×0.5 multiplier)
  • Temperature is outside 35–40 °C (×0.7)
  • SpO₂ < 85% (×0.6)
  • HR variance over 5 samples > 400 bpm² (×0.7)

apply_coherence_modulation then scales the desired infusion rate by coherence_sigma, so noisy or degraded telemetry reduces — rather than amplifies — infusion aggressiveness.


7. Configuration Constants and Bounds

All constants are centralised in src/config_defaults.hpp.

Control Timing

Constant Value Description
CONTROL_FREQUENCY_HZ 5 Control loop rate
CONTROL_PERIOD_SEC 0.2 s Period between control ticks

Infusion Rate Bounds

Constant Value Description
MIN_INFUSION_RATE_ML_MIN 0.1 ml/min Hard lower bound
MAX_INFUSION_RATE_ML_MIN 1.5 ml/min Hard upper bound

Risk Thresholds

Constant Value Description
LOW_RISK_MAX 0.3 Risk scores ≤ 0.3 are low-risk
MODERATE_RISK_MAX 0.6 Risk scores ≤ 0.6 are moderate
HIGH_RISK_THRESHOLD 0.75 Risk scores > 0.75 trigger throttling

Cardiac Safety

Constant Value Description
MIN_CARDIAC_RESERVE 0.2 Below this: LOW_CARDIAC_RESERVE warning
CARDIAC_LIMIT_THRESHOLD 0.3 Below this: controller applies sigmoid scaling

Rate-of-Change Safety

Constant Value Description
MAX_RATE_CHANGE_ML_MIN 0.3 ml/min Maximum step change per cycle

Volume Safety

Constant Value Description
DAILY_VOLUME_PER_KG_ML 35.0 ml/kg Base 24-hour volume limit

Prediction

Constant Value Description
PREDICTION_HORIZON_MIN 10 minutes Look-ahead window
UNCERTAINTY_GROWTH_PER_MIN 0.05 per minute Confidence decay in predictions

8. Failure Mode Analysis

Sensor Noise or Degraded Signal Quality

Behaviour: calculate_coherence reduces coherence_sigma. apply_coherence_modulation multiplies the desired rate by this value, reducing aggressiveness proportionally.

Result: Conservative, dampened response. Infusion does not increase on unreliable data.


Out-of-Range Heart Rate

Behaviour: HR outside 40–180 bpm halves the coherence score. If HR exceeds 1.4 × baseline_hr_bpm, TACHYCARDIA_DETECTED is raised and rate is capped at 0.4 ml/min.

Result: Dual protection: coherence damping reduces desired rate; safety monitor caps it.


Low SpO₂

Behaviour: SpO₂ < 85% reduces coherence by 40%. SpO₂ < 90% reduces tissue efficiency in the energy transfer calculation. SpO₂ < 95% reduces cardiac reserve via the reserve formula.

Result: Multi-layered response — energy proxy, cardiac reserve, and coherence all degrade, driving a conservative control trajectory.


Cumulative Volume Approaching Limit

Behaviour: update_volume accumulates delivered volume. When projected volume exceeds 90% of the daily limit, VOLUME_LIMIT_APPROACH fires and rate is capped at 0.3 ml/min.

Result: Throttling begins before the hard limit is reached.


Zero Cardiac Reserve

Behaviour: apply_cardiac_limiting applies a sigmoid scaling factor when cardiac_reserve < 0.3. Below 0.2, the safety monitor adds LOW_CARDIAC_RESERVE and caps at 0.5 ml/min. Emergency minimum rate (0.1 ml/min) is preserved if the patient is critically dehydrated.

Result: Layered protection; rate is reduced but life-sustaining minimum is maintained in extremis.


Prediction Unavailable

Behaviour: predict_forward requires at least 5 historical state samples. If fewer are available, it returns std::nullopt. The controller skips the predictive boost when no prediction is available.

Result: System operates on current state alone — conservative and safe.


9. Known Gaps and Limitations

The following gaps are acknowledged in the v4.2.0 codebase. They do not affect the validity of the current pre-clinical research scope but are relevant for future development.

Gap Detail Status
AdaptiveController has no unit tests Controller logic (calculate_base_rate, apply_coherence_modulation, etc.) is exercised only via integration. Deferred to v4.3
SystemLogger has no unit tests NDJSON output format, escape logic, and file I/O are untested in isolation. Deferred to v4.3
RestApiServer has no unit tests JSON serialisation and HTTP response formatting are untested. Deferred to v4.3
simulation_engine.cpp not in Makefile The SimulationEngine compiles but has no test target and is not exercised by CI. Known issue
PatientProfile has no default initialisers Partially initialised PatientProfile objects risk undefined behaviour from uninitialised reads in test code. Known issue
REST API binds to 0.0.0.0 by default No authentication, no TLS, and global network binding are unsafe outside an isolated research network. Research-only
REST API is single-threaded A slow or malformed HTTP connection will stall all subsequent requests. Acceptable for research use
Whitepaper simulation results not reproducible from this repository The multi-patient statistical results cited in the whitepaper require a validated simulation harness not included in this repository. Documented

10. Path to Clinical Validation

This document does not constitute clinical validation. The following steps are required before this system could be considered for regulated clinical use:

  1. Hardware-in-the-loop testing with certified infusion pumps and sensors
  2. Formal hazard analysis (e.g., FMEA, FTA) covering all failure modes
  3. Formal verification of safety-critical invariants (e.g., via TLA+, CBMC, or Frama-C)
  4. Controlled clinical trials with qualified medical oversight
  5. Regulatory submission — FDA Class II/III clearance or CE marking under MDR 2017/745
  6. Medical-grade cybersecurity hardening including authentication, TLS, and audit logging
  7. EHR and clinical workflow integration
  8. Expanded test coverage — at minimum, AdaptiveController, SystemLogger, and end-to-end simulation harness

This system is designed as a transparent, auditable research baseline that supports — but does not replace — those processes.